Thermo Scientific FlashSmart CHNS/O Elemental Analyzer
Description
The Thermo Scientific FlashSmart CHNS/O Elemental Analyzer is equipped with two independent furnaces each connected with a separate analytical circuit. The switch between analysis can be done sequentially and are completely automated through the MultiValve Control (MVC) Module. Carbon, nitrogen, hydrogen, and sulfur determination by combustion analysis and oxygen determination by pyrolysis are enabled with the current setting for the characterization of various solid materials in pharmaceutical, cosmetics, and other material research fields for quality control and R&D purposes. The Thermo Scientific FlashSmart Elemental Analyzer is the optimal solution for the analysis of CHNS/O in terms of accuracy, reproducibility, automation, speed of analysis, and cost per analysis.
Measurement Modes
Combustion or Pyrolysis
Sample Pans
Tin (CHNS Measurements) and Silver (Oxygen Measurement)
Instrument Features
Automated Operation. Easy Determination of CHNS/O. MultiValve Control Module. Thermal Conductivity Detector. Analysis from ppm to 100%. Dual Furnaces. Separate GC Columns. Four Certified Standards. Two Solid Autosamplers. High Sample Throughput.
Popular Applications
Perform Accurate Element Analysis of Five Elements (C, H, N, S, and O) in the following fields:
Organic Chemistry & Pharmaceuticals. Petro-chemistry & Energy. Environmental. Agronomy & Marine Science. Material Characterization. Food Safety.
Instrument Specifications
Detector: Thermal Conductivity Detector (TCD)
Carrier Gas: Helium (40 psi)
Carrier Gas Flow Rate: 100 to 140 mL/min
Standby Gas: Nitrogen (40 psi)
Standby Gas Flow Rate: 10 mL/min
Combustion Gas: Oxygen (45 psi)
Combustion Gas Flow Rate: 250 mL/min
Furnace Temperature (Operation): 950 °C (Combustion, Left), 1060 °C (Pyrolysis, Right)
Furnace Temperature (Standby): 476 °C (Combustion, Left), 531 °C (Pyrolysis, Right)
Sample Size: 0.5 to 2 mg
Autosampler: 32 Positions Each