Thermo Scientific FlashSmart CHNS/O Elemental Analyzer

Description

The Thermo Scientific FlashSmart CHNS/O Elemental Analyzer is equipped with two independent furnaces each connected with a separate analytical circuit. The switch between analysis can be done sequentially and are completely automated through the MultiValve Control (MVC) Module. Carbon, nitrogen, hydrogen, and sulfur determination by combustion analysis and oxygen determination by pyrolysis are enabled with the current setting for the characterization of various solid materials in pharmaceutical, cosmetics, and other material research fields for quality control and R&D purposes. The Thermo Scientific FlashSmart Elemental Analyzer is the optimal solution for the analysis of CHNS/O in terms of accuracy, reproducibility, automation, speed of analysis, and cost per analysis.

Measurement Modes

Combustion or Pyrolysis

Sample Pans

Tin (CHNS Measurements) and Silver (Oxygen Measurement)

Instrument Features

Automated Operation. Easy Determination of CHNS/O. MultiValve Control Module. Thermal Conductivity Detector. Analysis from ppm to 100%. Dual Furnaces. Separate GC Columns. Four Certified Standards. Two Solid Autosamplers. High Sample Throughput.

Popular Applications

Perform Accurate Element Analysis of Five Elements (C, H, N, S, and O) in the following fields:

Organic Chemistry & Pharmaceuticals. Petro-chemistry & Energy. Environmental. Agronomy & Marine Science. Material Characterization. Food Safety.

Instrument Specifications

Detector: Thermal Conductivity Detector (TCD)
Carrier Gas: Helium (40 psi)
Carrier Gas Flow Rate: 100 to 140 mL/min
Standby Gas: Nitrogen (40 psi)
Standby Gas Flow Rate: 10 mL/min
Combustion Gas: Oxygen (45 psi)
Combustion Gas Flow Rate: 250 mL/min
Furnace Temperature (Operation): 950 °C (Combustion, Left), 1060 °C (Pyrolysis, Right)
Furnace Temperature (Standby): 476 °C (Combustion, Left), 531 °C (Pyrolysis, Right)
Sample Size: 0.5 to 2 mg
Autosampler: 32 Positions Each

Vendor Contact

CE Elantech, Inc.
170 Oberlin Ave. N., Suite 5
Lakewood, NJ 08701

Sales: sales@ceelantech.com
Service: service@ceelantech.com
Phone: 888-CE-CHNSO (888-232-4676)
Fax: 732-370-3888