TA Discovery Series HR-2 Hybrid Rheometer
Description
The Discovery series Hybrid Rheometer (HR-2) from TA Instruments offers precise measurements of both viscosity and viscoelasticity of various materials, e.g., high viscosity fluids, pastes, polymer solutions and melts, and soft solids and as a function of temperature. Typical applications include viscosity measurements of LED encapsulants, solder pastes, fluxes, underfill, die-attach adhesives, glob top encapsulants, etc. for the electronic packaging industry; curing profiles of thermoset plastics and UV-curing adhesives and encapsulants, and rheology and viscoelasticity of polymers.
Available Geometries
Parallel Plate Geometries (20 mm and 40 mm)
Available Lower Plate
Standard Peltier Plate
What Can be Measured?
Low/Medium/High Viscosity Liquids, Complex Fluids, Structured formulations, and Soft Solids/gels
Sample Dimension Requirements
Materials with low modulus and high plasticity can be accepted as bulk. Materials with higher elasticity should be provided as disk with 20 mm diameter and thickness <2 mm.
Instrument Features
Smart Swap™ Geometries. Smart Swap™ Temperature Systems. Ultra-low Inertia Drag Cup Motor. Radial Air Bearings. Normal Force Rebalance Transducer (FRT). Direct Strain Oscillation. Second Generation Magnetic Thrust Bearing. New True Position Sensor (TPS). Ultra-low Compliance Single-Piece Frame. High-Resolution Optical Encoder. Nano-Torque Motor Control. Traceable Torque Calibration.
Popular Applications
Viscosity and Complex Viscosity Measurements. Shear-Thinning or Shear-Thickening Behavior. Thixotropic and Rheopectic Properties. Yield Stress measurement. Temperature-Dependent Viscosity/Modulus. Frequency-Dependent Viscosity/Modulus. Creep and Stress Relaxation Behaviors. Viscoelastic Properties of Materials.
Instrument Specifications
Bearing Type, Thrust: Magnetic
Bearing Type, Radial: Porous Carbon
Motor Design: Drag Cup
Minimum Torque Oscillation: 2 nN·m
Minimum Torque Steady Shear: 10 nN·m
Maximum Torque: 200 mN·m
Torque Resolution: 0.1 nN·m
Minimum Frequency: 1.0 × 10-7 Hz
Maximum Frequency: 100 Hz
Minimum Angular Velocity: 0 rad/s
Maximum Angular Velocity: 300 rad/s
Displacement Transducer: Optical Encoder
Displacement Resolution: 10 nrad
Step Time, Strain: 15 ms
Step Time, Rate: 5 ms
Normal/Axial Force Transducer: Force Rebalance Transducer
Maximum Normal Force: 50 N
Normal Force Sensitivity: 0.005 N
Normal Force Resolution: 0.5 mN
Instruction & Brochure
Vendor Contact
TA Instruments
109 Lukens Dr.
New Castle, DE 19720
Phone: 302-427-4000
Fax: 302-427-4041
Email: answerman@tainst.com
Website: http://www.tainstruments.com
Application Hotline: 302-427-4070
SOP & Knowledgebase
TEMPR Lab maintains an ever-growing knowledge base relevant to all techniques involved in the lab. This information together with instrument SOPs is made available to all trained users through shared Google Drive.