Thermo Scientific iCAP™ RQ Inductively Coupled Plasma Mass Spectrometer (ICP-MS) with Teledyne Laser Ablation (LA) System

Description

Inductively coupled plasma mass spectrometry (ICP-MS) is an elemental analysis technology capable of detecting most of the periodic table of elements at milligram to nanogram levels per liter. It is used in a variety of industries including, but not limited to, environmental monitoring, geochemical analysis, metallurgy, pharmaceutical analysis, and clinical research.

The Thermo Scientific™ iCAP™ RQ single quadrupole (SQ) ICP-MS combines high accuracy, low detection limits, and streamlined measurement protocol with a much-simplified software user interface. The Qtegra™ ISDS Software manages entire workflows from ‘Get Ready’ to data reporting. The autosampler enables 24/7 sample analysis for advanced applications in research.

The laser ablation solid sample introduction system from Teledyne includes a 193 nm laser for application in most fields. it allows for direct solid sample measurements bypassing the potential burden and complication when performing sample digestion. Elemental surface profiling is also available for 2D visualization.

Acceptable Sample Forms

Solution Introduction ICP-MS: Solution Sample in 1% Nitric Acid

Laser Introduction ICP-MS: Thin-film samples (with smooth surfaces) on microscope slides

Instrument Features

Compact Vertical Design. QCell with Flatapole Design. Helium Kinetic Energy Discrimination. Solid-State RF Generator. Self-Aligning Sample Introduction System. Drop-Down Door. Quadrupole Mass Filter. Solution Sample Autosampler. Laser Ablation Solid Sample Introduction. Qtegra Intelligent Scientific Data Solution.

Popular Applications

Automated elemental analysis applicable to a growing range of applications:

Environmental. Food Safety. Water Quality. Pharmaceuticals. Clinical Research. Materials Analysis. Geoscience Exploration. Biomedical Research. Forensic Science. Toxicology.

Instrument Specifications

Nebulizer: Concentric Borosilicate Glass with 0.4 mL/min Flow Rate
Spray Chamber: Baffled Cyclonic, High Purity Quartz
Torch: Quartz, Demountable
Injector: Quartz, 2.5 mm ID
Sample Cone: Ni, 1.1 mm Diameter Orifice
Skimmer Cone: Ni, 0.5 mm Diameter Orifice
Skimmer Insert: High Matrix Insert
Extraction Lens: Low Voltage Conical
Ion Optics: 90° RAPID Ion Lens
QCell: Collision Gas He (99.999%)
Mass Range: 2 to 290 u
Scan Speed: >3700 amu/s; Li to U with 40 Interval Masses
Mass Stability: <±0.025 u per day
Abundance Sensitivity: <0.5 ppm at m-1 (m = 238U)
Minimum Dwell Time: 100 μs in Pulse and Analog
Dynamic Range: >9 orders of magnitude (<1 to >1 × 109 cps)

STD Mode
Sensitivity: (kcps/ppb) 55 (7Li), 100 (59Co), 240 (115In), 330 (238U)
Detection Limit: (ppt) <0.5 (9Be), <0.1 (115In), <0.1 (209Bi)
Oxides (%): <2 (CeO/Ce)
Double Charged (%): <3 (Ba++/Ba+)
Background (cps): <1 (m/z 4.5)
Stability (% RSD): <2 (10 min), <3 (2 hours)
Isotope Ratio Precision (% RSD): <0.1 (107Ag/109Ag)

He KED Mode
Sensitivity (kcps/ppb): 30 (59Co)
Background (cps): <0.5 (m/z 4.5)