TA Instruments TGA Q500

Description

The TA Instruments Thermogravimetric Analyzer (TGA) Q500 is a thermal weight-change analysis instrument, used in conjunction with a TA Instruments thermal analysis controller and associated software, to make up a thermal analysis system. The Thermogravimetric Analyzer measures the amount and rate of weight change in a material, either as a function of increasing temperature or isothermally as a function of time, in a controlled atmosphere. It can be used to characterize any material that exhibits a weight change and to detect phase changes due to decomposition, oxidation, or dehydration. This information helps identify the percent weight change and correlate chemical structure, processing, and end-use performance.

The TA TGA Q500 has six major components: (1) The balance, which provides precise measurement of the sample weight. The balance is the key to the TGA system; (2) The sample platform, which loads and unloads the sample to and from the balance; (3) The furnace, which controls the sample atmosphere and temperature; (4) The cabinet, where the system electronics and mechanics are housed; (5) The heat exchanger, which dissipates heat from the furnace; (6) Two mass flow controllers, which control the purge gas to the balance and furnace.

Available Purge Gas

Nitrogen

Sample Pans

100 μL Platinum or Ceramic

Instrument Features

Low-Mass Furnace. Proprietary Heater Control Technology. Dual Digital Mass Flow Controllers. 16-Position Autosampler. Vertical Thermobalance. Horizontal Purge Gas System.

Popular Applications

Thermal Stability. Decomposition Kinetics. Compositional Analysis. Estimated Lifetime. Oxidative Stability. Moisture and Volatile Content Analysis.

Instrument Specifications

Maximum Sample Weight: 1 g
Weighing Precision: ±0.01%
Sensitivity: 0.1 μg
Baseline Dynamic Drift: < 50 μg
Furnace Heating: Resistance Wound
Temperature Range: Ambient to 1000 °C
Isothermal Temperature Accuracy: ±1 °C
Isothermal Temperature Precision: ±0.1 °C
Controlled heating Rate: 0.01 to 100 °C/min
Furnace Cooling: 12 min (1000 to 50 °C, Forced Air)
Temperature Calibration: Curie Point
Autosampler: 16 Positions